Digital Force Gauge Model:HF series
Force gauge is also called push pull gauge, adopt strain gague type load cell, digital gauge can do the pull test, measuring force, a force mesurement device.
Electronic digital force gauge can measure the force on compression and tension.
Widely used on connector insertion and other forces, such as, fracture, bend, peeling off, breakage, hardness, elasticity and so on.
The unit display is available for N, kgf, g. & lb. Variety capacity is opened for option.
Characteristics
1. STATISICAL PROCESSING & PRINTING OF INSPECTION REPORT
The HF series can calculate figures of the standard deviation and the process capability index and those figures can be printed out on your printer in form of the inspection report when the CENTRONICS INTERFACE(OPTIONAL) is adopted.
2. COMPARATOR FUNCTION
Initial input of the tolerance limit (upper/lower) enables the device to judge the measured result as GO/NO.
3. MEMORY FUNCTION
Measured results will be automatically memorized up to 256 measurements.
4. HIGH ACCURACY, HIGH RESOLUTION
The HF series have achieved extremely high accuracy; such as ±0.2%F.S. of accuracy and ±0.1%F.S. of repeatable accuracy. And also, the HF series perform extremely high resolution. For example, the 1kgf capable device can indicate minimum 0.1gf. It means 1/10000 resolution.
5. MULTIPLE FUNCTIONS
Initial setting on the device is possible for measurement of peak force, insertion/withdraw force, insertion compression peak force, withdraw extension peak force and holding force at contact point Also, the devices is applicable for setting of various printing modes, input of local Newton value and other various modes settings.
6. UNIT CONVERSION
It is possible to convert the graduations of N, kgf(gf) or lbf on the device.
7. PEAK HOLD FUNCTION
Single operation can perform measurement of both compression peak data at insertion and extension peak data at withdraw.
8. EASY OPERATION
Buttons are well arranged for easy operation th perform multipurpose measurement.
9. AUTOMATIC POWER OFF
The HF series have automatic power-off function to save battery and to keep the device safe. Power off time can changed at the range of 01-99min.
10. LOCAL NEWTON VALUE INPUT
The terrestrial gravitation varies with where measurement is performed. Measurement error by gravity difference will be eliminated by inputting the local Newton value according to the measurement place.
11. RS232C SETTING FUNCTION
It is possible to set data bit length, baud rate, stop bit length and parity bit to connect with your computer.
12. VARIOUS OUTPUT
Various options are offered for outputs; CENTRONICE output, RS-232C output, analog output, digimatic output and switch output.
Specifications
MEASUREMENT UNIT |
N, Kgf, (gf), lbf |
ACCURCY |
±0.2% F.S. |
REPEATABILITY |
±0.1% F.S. |
RESPONSE SPEED |
10~320HZ, 6 Steps |
DISPLAY |
4DIGIT LCD DISPLAY |
MEASUREMENT |
MAXIMUM VALUE•MOMENT VALUE |
SAMPLING |
640 TIMES PER SEC. |
ALLOWABLE LOAD |
150% (ALARM BUZZER AT 105%) |
A/D CONVERTER |
A/D CONVERTER 16bit |
OPERATION TIME |
APPROX.10 HOURS |
(FULL CHARGED AT 25℃) |
|
CHARGING TIME |
APPROX.6 HOURS |
OPERTION TEMPERATURE |
0~+40℃ |
POWER |
NI-MH BATTERY, |
A/C ADAPTOR FOR CHARGING (110V,220V) |
|
WEIGHT |
500g |
DIMENSIONS |
230x65x35mm |
STANDARD ACCESSARIES |
ATTACHMENT, A/C ADAPTER, CABLES, CASE |
OUTPUT |
STANDARDANALOG OUTPUT ±2V F.S. |
OPTIONALPRINTER CONNECTING, DIGIMATIC OUTPUT, |
|
CENTRONICS INTERFACE. COMPARATOR CONNECTING, |
|
RS-232 INTERFACE. |
Unit |
Max.Load |
Min.Resolution |
||||||
Model |
N |
Kgf |
gf |
lbf |
N |
Kgf |
gf |
lbf |
HF-1 |
10 |
1 |
1000 |
2.204 |
0.001 |
0.001 |
0.1 |
0.001 |
HF-2 |
20 |
2 |
2000 |
4.408 |
0.01 |
0.001 |
1 |
0.001 |
HF-5 |
50 |
5 |
5000 |
11.02 |
0.01 |
0.001 |
1 |
0.01 |
HF-10 |
100 |
10 |
10000 |
22.04 |
0.01 |
0.001 |
1 |
0.01 |
HF-20 |
200 |
20 |
… |
44.08 |
0.1 |
0.01 |
… |
0.01 |
HF-50 |
500 |
50 |
… |
110.2 |
0.1 |
0.01 |
… |
0.1 |
HF-100 |
1000 |
100 |
… |
220.4 |
0.1 |
0.01 |
… |
0.1 |
When display value is bigger than 999 on HF-1, HF-10, and HF-100, the 1000th digit will exchange between 1 and 0. It means the 1000th digit display value is 10000.